![horiba Fluorolog-3 Operation Manual Download Page 204](http://html1.mh-extra.com/html/horiba/fluorolog-3/fluorolog-3_operation-manual_2162215204.webp)
Fluorolog-3 v. 2.2 (9 Aug 2002)
TRIAX operation with the Fluorolog
®
-3
14-27
f
Click
CCD Regions
.
The
Acquisition
Setup
window opens.
g
Choose the
Position
index-card
tab.
Five different spectra
are required. The fol-
lowing example uses
a CCD that is 2000
pixels wide × 800
pixels high, with a
Hg line at 546.074
nm.
h
Set
Nb
Region
to 1.
i
Set the
Center
to 546.074 nm, and press
Enter
.
j
Calculate the increment
i
=
−
To From
5
, to give 5 positions using the
maximum amount of the chip in linearization.
k
Set
Nb Regions
to 5.
l
Set the first region’s
From
= 1.05 ×
i
.
m
Set the second region’s
Center
2
=
Center of region
1 +
i
.
n
Repeat steps 12 and 13 until all 5 regions are assigned.
o
Click
OK
.
The
Acquisition Setup
window closes. In the
CCD Acquisition
dialog box,
p
Enter a file name.