System Operation
3-16
Emission recalibration
Calibration for the Fluorolog
®
-3 is performed by:
•
Moving to the position of the observed peak,
•
Going to the
Visual Instrument Setup
application, and
•
Telling the software the correct position for this peak.
The software will save this change in position to the calibration file.
1
Note where the 397-nm peak was observed.
2
Open the
Real Time
Display
.
3
Reset the emission
monochromator to the
position where the peak
was observed.
Hit
Tab
on the keyboard to set the
monochromator to the entered
value.
4
Close the
Real Time
Display
.