Fluorolog-3 v. 2.2 (31 Jul 2002)
Producing Correction Factors
11-3
Generation
1
Open the
Real Time
Display
.
2
Turn off the high
voltage to the
emission
photomultiplier tube.
3
Close the slits.
4
Place the 1908MOD
Scatter Block
Assembly in the
sample chamber, so that
light is directed toward the
right angle.
Looking down, the scatter plate should be
toward the left.
5
Place the mask over the
sample compartment.
The square hole should be vertically
centered over the white scatter plate. Secure
the mask with black tape.