![horiba Fluorolog-3 Operation Manual Download Page 170](http://html1.mh-extra.com/html/horiba/fluorolog-3/fluorolog-3_operation-manual_2162215170.webp)
Fluorolog-3 v. 2.2 (11 Jul 2002)
Determining the Plateau Voltage
12-6
From these graphs, at ~1000 V there is a plateau both in the signal and dark
counts. Therefore, the point at which there is maximum voltage with minimum
dark counts is 1000 V—the plateau voltage for this detector.