Fluorolog-3 v. 2.2 (31 Jul 2002)
Producing Correction Factors
11-10
Select
S
(signal) detector.
Click
OK
to close this
box.
Note:
Remember to multiply each irradiance value by the wave-
length at which it is valid
prior
to attempting to enter its value. The
units will be photons·s
–1
·cm
–2
·nm.
a
Click
Signals...
to
open the
Signals
dialog box:
b
Click
HV
to
open the
High
Voltage
dialog
box:
By turning off the high voltage, a file which
has 0 intensity for each point will be
created. The irradiance values will be
entered for each 0 point.
5
Run the scan with these parameters.
6
Name the file
CRE
.
After performing the irradiance value multiplication and displaying the file
CRE
on the screen, enter each value as follows:
7
Select View from the toolbar.
8
Under View, choose Table View.
A table appears, displaying the wavelengths and the value of the intensity at
each wavelength. The table may be displayed showing only the last value at the
last wavelength. Click on the value (the number will become highlighted), and
use the arrow keys to scroll up and down.
Turn off the volt-
age to the detec-
tors.