![horiba Fluorolog-3 Operation Manual Download Page 127](http://html1.mh-extra.com/html/horiba/fluorolog-3/fluorolog-3_operation-manual_2162215127.webp)
Fluorolog-3 v. 2.2 (31 Jul 2002)
Troubleshooting
7-4
The following lamp scan spectrum shows poor resolution in the area around the peak.
Lamp scan of 150-W Xe lamp. Note poor resolution in the area near the
467-nm peak.
This lack of spectral resolution appears because the slit widths are set too wide. To re-
solve this problem, narrow the slit widths.
Intensity (cou
nts s
–1
)
0.4473
0.3355
0.2236
0.1118
0
Xenon-lamp peaks
are unresolved
300
400
500 600 700 800
Wavelength (nm)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com