![horiba Fluorolog-3 Operation Manual Download Page 159](http://html1.mh-extra.com/html/horiba/fluorolog-3/fluorolog-3_operation-manual_2162215159.webp)
Fluorolog-3 v. 2.2 (31 Jul 2002)
Producing Correction Factors
11-14
The correction factors shown above were acquired for a Fluorolog
®
-3 system
with 500-nm blazed gratings in the emission spectrometer and a red-sensitive
R928P photomultiplier detector.
Once the emission correction factors have been found, determination of the excitation
correction factors may be necessary. The following procedures describe how to obtain
excitation correction factors using the photomultiplier and the photodiode. Follow the
procedure that applies to your configuration.