![horiba Fluorolog-3 Operation Manual Download Page 13](http://html1.mh-extra.com/html/horiba/fluorolog-3/fluorolog-3_operation-manual_2162215013.webp)
Fluorolog-3 v. 2.2 (11 Jul 2002)
Requirements & Installation
1-3
Note:
For the computer, Jobin
Yvon Inc. recommends using a
surge suppressor or an
uninterruptible power supply
(UPS) with a surge suppressor.
Warning:
A three-
prong-to-two-
prong adapter
is
not recommended.
Electrical requirements
•
115 V, 20 A or 220 V, 20 A; factory-set.
•
As an extra measure of
caution, plug the xenon lamp
into a circuit separate from
the other components. This
guarantees that the electrical
surge from the lamp never
will interfere with the
computer or system.
Make sure enough AC outlets are available for the
•
Computer
•
Printer (optional)
•
Monitor
•
Xenon lamp
•
System controller (SAC)
Use three-prong plugs for proper grounding of the
system. If a two-prong adapter is used, for the
safety of the operator and to preserve the integrity
of the system, the adapter must be attached to the
wall outlet properly, according to the
manufacturer’s instructions. This provides a
positive connection to the electrical ground (earth),
ensuring that any stray or leakage current is
directed to earth ground.