Fluorolog-3 v. 2.2 (11 Jul 2002)
System Maintenance
5-17
Warning:
A circuit board and critical
connections are attached to the lid of
the spectrometer. DO NOT attempt to
lift the lid completely off.
Remove the existing grating
1
Close the slits in the emission spectrometer.
2
Turn off all instrument power.
3
Remove the Phillips-head screws holding the lid
on the
monochromator.
4
Lift the lid slowly
until you feel
resistance.
This is the maximum length of the cables attached to the lid.
5
Insert the lid support at the rear of the
monochromator.
6
Place the lid in the support.
Overhead view of the grating assembly.
Grating motor
Thumbscrew
Grating stand
Grating
Mounting pin