Fluorolog-3 v. 2.2 (11 Jul 2002)
System Operation
3-9
Excitation monochromator recalibration
Re-calibration of the Fluorolog
®
-3 is performed by moving to the position of the observed
peak, going into
Visual Instrument Setup
, and telling the software the correct position at
which this peak should be. The software will save this change in position to the calibration
file.
1
Note the wavelength where the 467-nm peak
was observed.
2
Open the
Real
Time Display
.
3
Set the excitation
monochromator to
the position where
the peak was
observed.
Hit the
Tab
key on the
keyboard to set the
monochromator to the
entered value.
4
Close the
Real
Time Display
.