![horiba Fluorolog-3 Operation Manual Download Page 69](http://html1.mh-extra.com/html/horiba/fluorolog-3/fluorolog-3_operation-manual_2162215069.webp)
Fluorolog-3 v. 2.2 (11 Jul 2002)
System Maintenance
5-2
Warning:
Do not remove the pro-
tective cover from the replace-
ment xenon lamp until in-
structed to do so.
Warning:
Xenon lamps are an explosion hazard. Make sure the power
is off and all AC power is disconnected from the system. Read and
follow the cautions presented below.
Replacement
The replacement xenon lamp is packed in
the manufacturer’s box and must be
installed in the lamp housing. Read all the
packing material including instructions and
precautions before attempting to insert the
lamp into the lamp housing.
Hazards
!
Xenon arc lamps are an explosion hazard. Wear goggles and protective clothing
when opening the lamp housing and when handling the lamp.
!
The lamp power supply should not be connected to an AC power line while handling
lamp leads. Lethal high voltages may exist.
!
The lamp will remain extremely hot for approximately one-half hour after it has been
turned off. Do not touch the lamp or the metal unit until the lamp has cooled.
!
Never look directly at the xenon arc or its reflection. Intense radiation can perma-
nently damage eyes.
!
Do not touch the focusing lens, back scatter mirror, or the surface of the lamp. Fin-
gerprints will be burned onto these surfaces when the lamp is ignited.
The following instructions are divided into three distinct procedures—each of which may be
performed as a stand-alone operation:
•
Open the lamp housing
•
Remove the existing xenon lamp
•
Install a new xenon lamp
!
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com