![horiba Fluorolog-3 Operation Manual Download Page 73](http://html1.mh-extra.com/html/horiba/fluorolog-3/fluorolog-3_operation-manual_2162215073.webp)
Fluorolog-3 v. 2.2 (11 Jul 2002)
System Maintenance
5-6
Note:
Make sure
the cables do
not block the exit
slit of the lamp
housing.
The positive end is marked with a “+”, and should be pointing upward.
5
Insert and tighten the 7/64" cap screw on the
upper bracket.
6
Secure the upper lamp lead and upper power
cable to the upper bracket,
using a 5/32" cap screw.
7
Secure the lower lamp lead
and lower power cable to the
lower bracket, using a 5/32"
cap screw.
8
Tighten these two 5/32" cap
screws on the upper and
lower brackets.
9
Pull the excess power cables away from the
lamp assembly beneath the baffle.
Diagram of lamp in place.
10
Replace the cover of the lamp housing and
secure with the ten Phillips-head screws.
Lamp lead
Power
cable
Do not block exit slit
Baffle
Lower bracket
Pull excess cable beneath internal
housing away from lamp assembly
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com