![horiba Fluorolog-3 Operation Manual Download Page 211](http://html1.mh-extra.com/html/horiba/fluorolog-3/fluorolog-3_operation-manual_2162215211.webp)
Fluorolog-3 v. 2.2 (9 Aug 2002)
TRIAX operation with the Fluorolog
®
-3
14-34
matically (
Areas Definitions
) and graphically (
Areas Views
).
Comments:
Unlimited amount of information may be entered, but only the first 80 charac-
ters are displayed.
Linearization
Normally, data
are taken as
pixel number
versus
intensity. The
Linearization
checkbox lets
the user
convert to
desired units
versus inten-
sity, applying
linearization
corrections.
Normalize
Visible only
with two or
more active
areas,
Normalize
lets
the user choose
an area as the
normalization standard. DataMax performs a
Dark Subtract
on all data, then
divides all areas by the normalization standard.
Reformat
Reformat
divides the entire CCD chip equally (to within 1 pixel) into the value
shown in the
Number of Areas
slider. If
Number of Areas
is 1, then the en-
tire CCD is used as one area. In
Image
mode, only one area is allowed.
Changes are shown in the
Areas Definitions
.
Load
and
Save
Load
and
Save
custom CCD-area descriptions for later use, with an
.arl
extension.
3
Adjust all parameters as desired.
4
Click OK.
The
Areas List
closes.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com