Fluorolog-3 v. 2.2 (11 Jul 2002)
System Operation
3-6
b
Click on
Signals...
This opens the
Signals
dialog box.
Enter
R
(reference), then
click
OK
to close the box.
c
Click on
Slits...
This opens the
Slits
dialog box.
Set the excitation
entrance and exit slits
to
0.5
mm.
Set the emission
entrance and exit slits
to
0
mm (closed).
Click
OK
to close the
window.
If the slits’ units are not
in millimeters (mm), then open
Visual Instrument Setup
, select
Options
,
then
Units
, and set the slits’ units to
mm
. Use the reference (
R
) detector
channel.
d
Click
Run
to execute the scan.