Optimizing Data Acquisition
4-7
Note:
If the signal level exceeds 2 × 10
6
cps, decrease the slit
widths by 50%. If you do not see an obvious peak, increase
the excitation wavelength and the starting and ending wave-
lengths by 25 nm, and acquire another emission scan.
11
Enter the data file name and experiment file
name, and any comments desired to distinguish
this experiment from others.
12
Click Run to execute the experiment.
13
With the preliminary emission spectrum on the
screen, note the greatest intensity.
14
Either
Record the wavelength at which this occurs. This is your emission
maximum.
Or
Repeat this procedure until you find an emission peak.