Fluorolog-3 v. 2.2 (9 Aug 2002)
TRIAX operation with the Fluorolog
®
-3
14-7
The
New Value
dialog box closes.
5
Run a scan with the grating, to find a known
peak.
Jobin Yvon Inc. suggests running a xenon-lamp scan to find the 467-nm peak,
or a water-Raman scan to find the 397-nm peak. This procedure is similar to
that given in Chapter 3.
6
Note the position of the found peak.
As an example, we assume a lamp scan was performed, and the 467-nm peak
was observed at 480 nm.
7
Open the
Real Time
Display
.
8
Enter the observed
peak position.
Don’t forget to hit the
Enter
key.
Here, we entered 480 nm, the
position where the 467-nm peak
was seen.
9
Close the
Real Time
Display
.
10
Click Calibrate….
The
Enter Correct
Position
dialog
box opens.
11
Enter the true peak
position.
12
Click OK.
The
Enter Correct Position
dialog box
disappears.