Fluorolog-3 v. 2.2 (10 Sep 2002)
Optimizing Data Acquisition
4-19
Note:
Before applying correction factors, Jobin Yvon
recommends subtracting the dark counts (because dark
counts are not wavelength-dependent), and the spec-
trum of the blank, from the data. Refer to the software
manual for specific instructions.
During acquisition
Data can be acquired either as raw data or as corrected data. A spectrum composed of raw
data exhibits the effects of system parameters, while a corrected spectrum displays only the
properties related to the sample. To automatically acquire corrected emission data, enter
mcorrect
in the
Correction
factor file field in the
Run Experiment
dialog box. To acquire
corrected excitation data, enter
xcorrect
.