Fluorolog-3 v. 2.2 (31 Jul 2002)
Troubleshooting
7-2
Communication prob-
lems between com-
puter and instrument.
Boot disk corrupted.
Use backup boot disk. If no backup boot disk
is available, call Fluorescence Service De-
partment.
Cables are improperly con-
nected.
Check communications cables’ connections.
Wrong settings for COM
ports.
Check
datamax\isa_ini\sac.ini
file.
Computer I/O controller or
SAC I/O controller is fail-
ing.
Replace I/O controller: Call Fluorescence
Service Department.
Hardware Init. error
appears.
Broken IR sensor in mono-
chromator.
Replace IR sensor: Call Fluorescence Service
Department
.
Error converting slit
units appears.
Datamax\isa_ini\df
lt.set
file is corrupt.
Delete
datamax\isa_ini\dflt.set
file. When system is restarted, the system will
recreate this file.
Sample turret is not
operating.
Not turned on.
Verify that the rocker switch on the front
panel of the Model 1976 Accessory Controller
is in the “1” (
On
) position.
Software is not enabled.
Check status.
Cables are improperly con-
nected.
Check cable connections.
Instrument Control
Center
icon is not on
desktop, but appears
in taskbar
isamain.ini
is improp-
erly set
In the
windows
directory, open
isamain.ini
Change the coordinates listed to
Left = 524
Top = 174
Right = 824
Bottom = 282
If problem persists, change
isamain.ini
’s
properties into a read-only file.
Error “no trace view
object” appears
isascan.set
or
isas-
can.vw
are corrupted
Delete
isascan.set
and
isascan.vw
from datamax directory. (They will be recre-
ated automatically.)