Fluorolog-3 v. 2.2 (11 Jul 2002)
System Operation
3-17
Note:
The
Visual
Instrument Setup
will vary, de-
pending on the
instrument con-
figuration.
5
Open the
Visual Instrument Setup
dialog box:
6
Click on the
grating for
the
emission
mono-
chromator.
The
Grating/Turret
dialog
box appears:
7
Click on the
Calibrate... button.
This opens the
Enter Correct
Position
dialog box:
8
Enter 397 nm, the
actual peak position
using 350-nm
excitation.
9
Click OK to close the dialog box.
The emission monochromator now should be calibrated.
10
Click OK to close the
Grating/Turret
dialog box.