Fluorolog-3 v. 2.2 (31 Jul 2002)
Producing Correction Factors
11-19
Find the minimum data point and divide the file by that value.
25
Save the normalized file as
xcorrect
.
This overwrites the existing excitation correction-factor file.
To acquire corrected data for an experiment, enter the name of this file in the
COR-
RECTION
factor file field in the
Data Acquisition Parameters
dialog box. The
COR-
RECTION
factor file must be in the same directory as the data to be acquired and cor-
rected.