Fluorolog-3 v. 2.2 (9 Aug 2002)
TRIAX operation with the Fluorolog
®
-3
14-3
TRIAX
Fluorolog
®
-3
sample
compartment
CCD
SpectrAcq
Host PC
CCD Controller
Baseplate adapter
RS-232 cable
GPIB cable
37-pin
cable
BNC
cable
Automated slits
The TRIAX has automated adjustable slits at the entrance (2 mm maximum) and exit (7
mm maximum) ports. The slit width may be adjusted in steps of 0.0125 mm. For a 0.6-
mm slit on the TRIAX 320, dispersion is 1.58 mm.
Connections to system
Below is a sketch of the special cable connections required between a TRIAX spec-
trometer, the CCD detector, and the system. Details of these connections are given on
the following pages. Other Fluorolog
®
-3 connections are not shown.
Attachment to sample compartment
Fluorolog
®
-3 users typically choose the
lateral entrance port as the path for lumines-
cence from the sample compartment into the
TRIAX. To attach the TRIAX to the sample
compartment, a baseplate adapter must be
inserted between the sample compartment
and the TRIAX.
Baseplate attached to lateral
entrance port.