Fluorolog-3 v. 2.2 (10 Sep 2002)
Optimizing Data Acquisition
4-11
Note:
For weak signals, enter the
G
factor, rather than measure it auto-
matically. This may improve the S/N.
Either enter the
G
factor or have it meas-
ured automatically during the scan.
Run Experiment
For general experimental use of polarizers, use
Run Experiment
. Be sure that a layout in-
cluding polarizers is loaded. When selecting the type of experiment, the
Polarization
checkbox must be active. A typical
Emission Acquisition
dialog box is as follows:
Note the
G Factor
field that appears.
• To
measure
the
G
factor automatically, set the field to 0.
• To use a pre-determined
G
factor, enter it here.