Fluorolog-3 v. 2.2 (31 Jul 2002)
Producing Correction Factors
11-18
Note:
Doing so replaces the
xcorrect
file
that was shipped with the system.
Note:
Be sure the Auto Zero function is on.
21
Click Run to execute the scan.
22
Save the file
as
xcorrect
.
23
If no zeroes are shown, invert the data.
In
Run Experiment
, use the
Arithmetic
menu to divide the spectrum into 1.
24
Normalize the data.
Start the scan at 240 nm.
End the scan at 600 nm.
Increment by 5 nm
after each data point.
Set the
Inte-
gration
Time
to 5 s.
Perform one scan.