Fluorolog-3 v. 2.2 (31 Jul 2002)
Troubleshooting
7-6
If the problem goes away, then the problem was due to the cuvette surface. Clean or use
a different cuvette.
Or
1
Clean the cuvette.
2
Fill with fresh, double-distilled, deionized water.
If the problem goes away, then the problem was due to contaminated water.
Light not striking cuvette
The following graph shows a normal water-Raman scan with a superimposed problem
scan.
Here the problem is low intensity of the water signal when compared with the superim-
posed typical water Raman scan. To resolve this problem:
1
Make sure the cuvette is filled to the proper
level.
Light should fall on the sample, and the meniscus should not be in the light
path.
2
Make sure that the excitation and emission slits
are set to the proper widths.
3
Verify that the detector is set to the proper
voltage.
Low intensity during a Raman scan.
0
1
2
3
4
5
365
380
395
410
425
440
455
Wavelength (nm)
Intensity (10
5
counts/s
)
Clean water
Low signal