Fluorolog-3 v. 2.2 (31 Jul 2002)
Producing Correction Factors
11-16
Note:
Obtain a signal intensity of
no greater than 1.5 × 10
6
cps.
The slit width discovered in
this step will be used to run the
scan.
9
With the
Real Time
Display
still running, open
Run Experiment
.
10
In
Run Experiment
, choose Collect.
11
Under Collect, choose Experiment.
This opens the
Emission Acquisition
dialog box:
12
Click Exp
Type.
This opens the
Select
Experiment Type
dialog box:
13
Choose
Excitation
Acquisition,
then OK.
This closes the
Select Experiment Type
window, and then resets the
Emission
Acquisition
dialog box to
Excitation Acquisition
.