Fluorolog-3 v. 2.2 (9 Aug 2002)
TRIAX operation with the Fluorolog
®
-3
14-23
In
Continuous
mode,
a
Choose the number of
Accumulations
.
b
Set the
Integration Time
.
c
Choose the
Time Increment
from the start of one exposure to the start
of the next exposure.
d
Set the
Enable Int. Shutter Move
checkbox. If checked, the CCD shut-
ter opens and closes at each accumulation. If unchecked, the CCD shut-
ter opens at the start of the experiment, and closes at the end.
e
Decide upon a
Last Read Out Time
.
3
Set all appropriate parameters.
4
Click Run.
The experiment begins.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com