Fluorolog-3 v. 2.2 (11 Jul 2002)
System Operation
3-5
converts to
Excitation Acquistion
. (For calibration and calibration
verification, always adjust the excitation spectrometer first.)
5
Set the scan parameters for the xenon lamp
scan:
a
In the
Excitation Acquisition
dialog box:
Name the experimental parameters
lamp.exp
,
Start the scan at 220 nm,
Step the wavelength by 0.5 nm,
Set the emission to 650 nm,
Perform 1 scan.
Use a filename
lamp.spc
,
End the scan at 600 nm,
Use an
Integration Time
of 0.1 s.
Describe the file as the spectral profile of the xenon lamp.