![horiba Fluorolog-3 Operation Manual Download Page 175](http://html1.mh-extra.com/html/horiba/fluorolog-3/fluorolog-3_operation-manual_2162215175.webp)
Fluorolog-3 v. 2.2 (11 Jul 2002)
Reassembly Instructions
13-5
Connecting the reference detector to the SAC
The Fluorolog
®
-3 spectrofluorometer system comes standard with a silicon-photodiode
reference detector inside the sample-compartment module. The connections to the ref-
erence detector are on the outside front of the sample compartment module. The sample
compartment module (reference detector) is joined with the SAC via a split cable; that
is, a cable with three ends.
1
Find the double-headed
end of the split cable
#33979A.
2
Plug the end with the 7-pin
D-shell connector into the
D-shell jack on the front of
the sample compartment
module. Plug the BNC plug
into the BNC jack next to
the D-shell jack.
3
Plug the remaining D-shell
end of the cable into the INPUT R connector on
the back panel of the SAC.
DM302 PC module and signal-detector connections
1
Plug the end of the high-voltage cable #34040
with the MHV connector into the signal detector.
2
Connect the other end to the HV1 SHV
connector on the SAC.
3
The DM302 PC module has three plugs labeled
SIGNAL OUT, POWER and IN. Plug the short
cable #30645 with the BNC connector from the
signal detector to the IN connector on the
DM302.
The DM302 is joined to the SAC via the split cable #33977.
D-shell connector
BNC connector