4.3.2.1 low energy Timeline Toolbar
The toolbar contains the following:
Icon
Description
Lock - The Lock button only appears in live mode and is automatically depressed when the
user scrolls.
Unlock
First Packet
Previous Packet
Next Packet
Last Packet
Previous Interframe Spacing (IFS) Error
l
Interframe Spacing is considered valid if it is within 150 μs + or – 2us
l
If the Interframe Spacing is less than 148 us or greater than 152 us but less than or equal to 300
μs, it is considered an IFS error.
Next Interframe Spacing (IFS) Error
l
Interframe Spacing is considered valid if it is within 150 μs + or – 2us
l
If the Interframe Spacing is less than 148 us or greater than 152 μs but less than or equal to 300
us, it is considered an IFS error.
Previous Error Packet
Next Error Packet
Zoom In
Zoom Out
Reset - The Reset button appears only in live mode. Reset causes all packet data up to that
point to be deleted from the Packet Timeline display. This does not affect the data in Frame
Display. Resetting the display may be useful when the most recent throughput values are of
interest.
Table 4.3 -
Bluetooth
low energy Timeline Toolbar
4.3.2.2 low energy Timeline Menu Bar
The
Bluetooth
low energy Timeline
menu bar contains the following:
TELEDYNE LECROY
Chapter 4 Capturing and Analyzing Data
74
Frontline BPA low energy Hardware & Software User Manual