Variables Available for the final Evaluation
Setting Up the Print Form Using PDM
FISCHERSCOPE
®
X-RAY
79
@ANBOrder number of
the last block of the selected blocks
or the information that has been entered into the first entry
field of the dialog window “Order No./ Operator”.
@BM1Lot number of
the last block of the selected blocks
or the information that has been entered into the second entry
field of the dialog window “Order No./ Operator”.
@EDBNumber of
the last block of the selected blocks
.
@EDTDate of
the last block of the selected blocks
.
@ETMTime of
the last block of the selected blocks
.
@HG1Histogram of all single readings of the selected blocks
in the current product.
See also @WN1.
Regarding the numerals
1; 2; 3 … after certain
variables:
A measurement consists
of one or more single
readings (“channels”) that
are identified by consecu-
tive numbers (1,2,3).
Example “Dual coating
measurement”
Top coating thickness = 1
Intermediate coating
thickness = 2
@OE1Upper control limit UCL(d) of the mean values of an
SPC chart.
@OG1Upper specification limit USL (upper tolerance) for the
presentation and computation of the upper limit
violations in the selected blocks of the current product.
@OR1Upper control limit UCL(s) or UCL(R) of the standard
deviation s or the range R of an SPC chart.
@P<1Percentage of single readings in the selected blocks of the
current product that fall below the lower specification limit
LSL.
@P>1Percentage of single readings in the selected blocks of the
current product that exceed the upper specification limit
USL.
@PRFOperator of the last block of the selected blocks
or the information entered into the third entry field of the
dialog window “Order No./ Operator”.
@RG1Range = Difference between the largest (maximum) and
smallest (minimum) single readings of the selected blocks
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......