138
FISCHERSCOPE
®
X-RAY
Programming Coordinates for Automatic Measurements (XDLM 237)
17.7 Programming Individual Points
You can program any desired points on the XY(Z) stage, travel to each individual point and then
store its coordinates.
Before you start
1. Open the window
XY(Z) coordinates,
ref. Chap. 17.5 on Page 136.
2. Delete all coordinates that are no longer applicable, ref. Chap. 17.6 on Page 137.
Procedure
1. From the programming field select the option
Points
.
2. Travel to the desired measurement point: In the video image, use the mouse to click the desired
location or use the joystick at the front of the instrument.
3. Click
Store
.
4. Repeat steps 2 to 3 for each desired measurement point.
What you can do next
You can now travel to all programmed points (ref. Chap. 17.13 on Page 143) or program one or
two datum points (ref. Chap. 17.12 on Page 142).
17.8 Programming a Line
To program a line, define the starting point and the end point and the number of interim points on
the line.
Before you start
1. Open the window
XY(Z) coordinates
(ref. Chap. 17.5 on Page 136).
2. Delete all coordinates (ref. Chap. 17.6 on Page 137).
Procedure
1. From the programming field select the option
Points
.
2. Travel to the starting point of the line: In the video image, use the mouse to click the desired
location or use the joystick at the front of the instrument.
3. Click
Store
.
4. In the field
No. of intermediate pos.
, enter the number of points on the line.
5. Travel to the end point of the line: In the video image, use the mouse to click the desired loca-
tion or use the joystick at the front of the instrument.
6. Click
Store
.
The line is generated from the starting point, the end point and the entered interim points. The
interim points are distributed evenly over the distance.
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......