FISCHERSCOPE
®
X-RAY
147
Calibration, Base Correction and Reference Measurement
FISCHER can supply the following calibration standards:
Before you start
The instrument must be at operating temperature before performing the calibration.
Perform the calibration at the same workplace as you will perform the measurements. If you set
up the instrument at another place, perform the calibration at the new place again.
To get the best results, start a measurement for 20 minutes by performing the following steps:
a Click
Display Spectrum
.
b Click
Go
.
c
Wait for 20 minutes.
d Click
Stop
.
e
Click
Exit
.
Base material of the cali-
bration standard set (same
as in the definition of the
measurement application,
Def.MA)
Required only if the base material is
not
a pure element.
Pure element calibration
standard(s)
Required only if the option
Measure elements
has been enabled in
the definition of the measuring application (Def.MA), i.e., the spec-
tra library is not used to compute the pure elements.
Materials to measure the
spectra background
Required only in special cases.
Calibration standard set
(Cal-NS)
Consists of a calibration standard set with several calibration
standards of a specific coating/base material combination.
Individual calibration
standard
(Cal-N)
Individual calibration standard of a certain coating/base material
combination
Calibration foil
(Cal-F)
Calibration foil consisting of a free carrying layer of a certain ele-
ment. Universally usable on many different base materials.
Customer-specific
standards
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......