Trueness and precision
Measurement device monitoring for the Fischerscope X-RAY
FISCHERSCOPE
®
X-RAY
115
This can be observed, for example, with round robin tests on the
same sample with a marked reference area, even if each
instrument is correctly calibrated and checked. The reasons are the
random deviations of each single reading, and thus of each mean
value of a test series, the random deviations at each calibration,
and also the fact that each calibration standard is labeled with a
nominal value that through its measurement is also afflicted with an
unavoidable deviation from the true value (expressed by the
uncertainty of the standard). The sum of these deviations is
expressed in the
comparative precision
of the measurement
method. The repeatability precision characterizes the random
deviations of the results that are obtained at
one
instrument by
one
person with one and the same sample at the same measurement
location at repeated measurements, i.e., under repeatability
conditions. Comparative precision describes the random deviations
between different instruments, calibrated with different calibration
standards, at repeated measurements performed by different
persons at the same sample using the same measurement
locations (measurements under comparable conditions).
Theoretically, there should be trust between the supplier and
customer. However, the reality is: ”Trusting is good, inspecting is
better!”. One of the most important prerequisites for the inspections
is correct measurement readings. However, different readings
between supplier and customer should generally not be taken as a
cause to continuously doubt the measurement results or
measurement methods of the other party or to demand arbitration
measurements with different methods that are to verify that one’s
own measurement is correct. First one should investigate whether
the measurement results have been obtained at the same
measurement spots and the same specimens. To answer this
question, supplier and customer should exchange and measure the
same samples with marked measurement spots. If the results are
still different, the involved instruments should be calibrated using
certified calibration standards. This determines which instrument
measures correctly and which instrument requires corrective
measures. If the differences remain and the problem cannot be
resolved, only a well-founded consultation through experts of the
instrument manufacturer may help in the end.
Each user who monitors the measurement devices as described
herein, can verify the trueness of his own measurements at any
time. He can also trust his measurement results and will not be
troubled by unwarranted complaints. In practical situations, the
systematic deviations, which almost always arise from inadequate
or infrequent monitoring of the measurement devices, occur more
significantly and more frequently and above all are of a greater
magnitude than the random deviations between the instruments of
different parties. Typically, the random deviations are hardly
noticed and if they are noticed, they are accepted due to their
insignificance. Thus, random deviations as opposed to systematic
ones can usually be neglected as the reason for measurement
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......