FISCHERSCOPE
®
X-RAY
211
Features of WinFTM BASIC and LIGHT
2.6
Automated Material Identification (Class of Materials)
You can easily sort specimens with different compositions and/or coating thicknesses in different
classes of materials. You don’t need to perform reference measurements for each composition or
coating thickness. Each class of materials covers a range of composition and/or coating thickness.
Example: You need to sort the following specimens:
5 – 8 μm Zn/Fe
10 – 20 μm Zn/Fe
5 – 12 μm NiZn/Fe
In WinFTM you define three classes of materials, perform one measurement and WinFTM displays,
what class of materials the specimen belongs to.
Additionally, you can assign a product to each class of materials. The exact composition is meas-
ured. This has the following advantages:
The right product containing the appropriate parameters and calibration is loaded automati-
cally.
Measurement errors, caused by wrong product selection, are avoided.
2.7
Measurement Methods
Measurement by x-ray emission
Measurement by x-ray absorption
Measurement by simultaneous x-ray emission and x-ray absorption (Ratio Method)
Mathematical deconvolution
Measurement by Fundamental Parameter Method
The measurement methods comply with the following standards:
Standard
Titel and Edition
ASTM B568
Standard Test Method for Measurement of Coating Thickness by
X-Ray Spectrometry, 98 (2004)
DIN EN ISO 3497
Metallic Coatings – Measurement of Coating Thickness – X-ray
Spectrometric Methods, 12 (2001)
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......