17.12 Programming Datum Points
In addition to the measurement points (individual points, lines, arrays or circular coordinates), you
can also program one or two datum points. You can then change parts, or place parts anew and
ensure that always the correct coordinates are visited, even if the specimen is not positioned 100%
at the same place.
You can program one or two datum points.
Program one datum point, if you can be certain that the placed parts cannot be turned,
ref. Fig. 17.2.
Figure 17.2:Working with one datum point (1: Datum point, X: Measurement points)
Program two datum points, if the parts can be turned as well when they are placed, ref. Fig. 17.3.
Figure 17.3:Working with two datum points (1,2: Datum points, X: Measurement points)
Before you start
1. Open the window
XY(Z) coordinates
(ref. Chap. 17.5 on Page 136).
2. Program the measurement points (individual points, lines, arrays or circular coordinates).
Procedure
1. From the programming field select the options
1 Datum Point
or
2 Datum Points
.
2. Travel to the first datum point: In the video image, use the mouse to click the desired location
or use the joystick at the front of the instrument.
3. Click
Dat.1
.
The datum point is stored.
4. If you work with only one datum point, skip the steps 5 and 6.
5. Travel to the second datum point: In the video image, use the mouse to click the desired location
or use the joystick at the front of the instrument.
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......