Measurement device monitoring for the Fischerscope X-RAY
Pre-run for determining the control limits
122
FISCHERSCOPE
®
X-RAY
control chart as lines and correspond to the limit values that lead to
a correction if they are violated during subsequent measurements.
The final protocol of the pre-run generated with the WinFTM-PDM
software at the actual reference sample Ni/Cu/Fe 10-245 shows
beneath the group numbers (designated with "n
bl
"), the group
results (mean values Ni. and Cu. per group, standard deviation s
per group), the final result, the control limits as well as the control
charts for the nickel coating and for the copper coating:
Helmut Fischer GmbH Office West
Danziger Straße 1050374 Erftstadt
Tel.: 02235 41447Fax: 02235 922990
Fischerscope
®
XRAY Final protocol
___________________________________________________
Product:
Ni/Cu/Fe reference specimen 10-245
Folder:
ISO9000
Meas. application: Ni/Cu/Fe
Meas. mode:Dual coating (dd_)
Order:Pre-run
Operation.:5 x 10 measurements
Final evaluation fromGroup No. 1 measured 12:22:09
toGroup No. 10 measured 12:39:55
nbl= 1 Ni.= 7.54
μ
m Cu.= 7.79
μ
m
s = 0.085
μ
m s = 0.098
μ
m
nbl= 2 Ni.= 7.45
μ
m Cu.= 7.70
μ
m
s = 0.053
μ
m s = 0.155
μ
m
nbl= 3 Ni.= 7.36
μ
m Cu.= 7.68
μ
m
s = 0.123
μ
m s = 0.097
μ
m
nbl= 4 Ni.= 7.43
μ
m Cu.= 7.57
μ
m
s = 0.078
μ
m s = 0.174
μ
m
nbl= 5 Ni.= 7.45
μ
m Cu.= 7.54
μ
m
s = 0.155
μ
m s = 0.093
μ
m
nbl= 6 Ni.= 7.39
μ
m Cu.= 7.61
μ
m
s = 0.171
μ
m s = 0.148
μ
m
nbl= 7 Ni.= 7.46
μ
m Cu.= 7.69
μ
m
s = 0.111
μ
m s = 0.160
μ
m
nbl= 8 Ni.= 7.35
μ
m Cu.= 7.66
μ
m
s = 0.103
μ
m s = 0.135
μ
m
nbl= 9 Ni.= 7.49
μ
m Cu.= 7.72
μ
m
s = 0.215
μ
m s = 0.168
μ
m
nbl= 10 Ni.= 7.56
μ
m Cu.= 7.84
μ
m
s = 0.158
μ
m s = 0.101
μ
m
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......