FISCHERSCOPE
®
X-RAY
219
Features of WinFTM BASIC and LIGHT
Calibration standard set
Input and storing of nominal values of the used calibration stand-
ards is performed in the
Calibration standard set
in the measure-
ment application of the product. After the calibration has been
performed, the
Calibration standard set
contains, in addition to the
nominal values, the calculated actual values including the correc-
tion. You can use this data to create a document for monitoring the
measuring equipment.
Calculating and saving
measurement parameters
and calibration results
Measurement parameters like measuring time, calibration time,
correction data, data of the last normalization and calibration, etc.
can be calculated and saved.
Library of pure element
spectra
The library of pure element spectra makes measurement of pure el-
ements unnecessary when performing a calibration. This is impor-
tant in cases, where pure elements are not available.
Spectra library
During the installation of the instrument the spectra of 12 or 14 el-
ements (depending on the detector type) are measured and saved
in the spectra library. Using these spectra all further spectra are
calculated by the instrument.
You can control whether the elements are measured or the spectra
of the spectra library are used during calibration.
Reference measurement
The Reference measurement checks and adjusts the drift of the in-
strument. It is also used for monitoring the measuring equipment.
Stability test of the tube and
of the detector
The instrument performs automatic long-term self tests to check the
primary radiation, the resolution and gain of the spectrometer (de-
tector and electronics).
Automatic reduction of the
x-ray tube current
After several minutes of instrument idle the x-ray tube current is re-
duced automatically. Thus, the service life of the tube is increased
and energy is saved.
Drift compensation
A shift of the peak position in the spectrum is compensated by the
software. This avoids wrong results especially when there are crit-
ical peak overlaps.
Function
Description
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......