User Interface of the WinFTM Software
Intensity: Dependency on the Measurement Distance
The total intensity of the X-ray fluorescence radiation emitted by the specimen decreases with the
distance D of the detector from the sample with
1/D
2
, because the solid angle of radiation collect-
ed by the detector decreases with an increase in the distance as well, while the area of the radia-
tion entry window of the detector remains the same.
For the coating thickness of a sample to be computed correctly, the distance D is automatically tak-
en into account by the evaluation software.