Solution Analysis
19.3 Positioning the Measuring Cell
Before you start
Before you position the measuring cell you need to fill it, see Chap. 19.2 on Page 156.
Procedure
1. Place the measuring cell into the measuring chamber. Depending on the instrument type, the
cup opening must face upwards or downwards:
•
Instrument with measuring direction from top to bottom: Cup opening must face upwards.
•
Instrument with measuring direction from bottom to top: Cup opening must face down-
wards.
To make the focusing easier, position the measuring cell such that the measurement beam does
not point to the center, but to the edge curvature of the measuring cell, see the following figure:
2. Adjust the focus of the instrument.
3. Shift the measuring cell carefully such that the measurement beam points to the center of the
measuring cell. Do not readjust the focus. The video image shows the metal reference plate out
of focus. This is correct. To measure correctly, the focus must always be set to the surface of
the foil.
What you can do next
You can now perform measurements.
5. Place the clamping ring over the foil-covered
measuring cell and press it down
carefully
until the foil is taught. Avoid:
•
Tearing the foil
•
Air bubbles
Overflowing liquid is caught in the drip
groove.
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......