170
FISCHERSCOPE
®
X-RAY
Measurement Data Export
Header export mask
Syntax:
@CAL@TAB#@TAB@PRN@CRX@LFX@LFX
#@TABOperator@TAB@EL1 @DM1@TAB@EL2 @DM2@TAB@EL3 @DM3 …
@EL20 @DM20@LFX@LFX@END
Table 20-6: Explanation of the elements
For more clarity the syntax is shown here in two lines. In the window Definition of Export
Template the template must be written in one line to avoid unwanted carriage returns and
line feeds.
Variable
Explanation
@CAL
Contains the product name.
@TAB
A tabulator will be inserted.
#
The character # will be displayed.
@PRN
Contains the product number.
@CRX
A carriage return will be inserted.
Operator
The text
Operator
will be inserted.
@LFX
A line feed will be inserted.
@EL
Contains the name of the result channel (element, coating thickness, ppm etc. ),
must always be supplemented with the number of the respective channel, e. g.
@EL1
@DM
Contains the unit of the result channel (mils, nm, % etc.), must always be supple-
mented with the number of the respective channel, e. g. @DM1
@END
Ends the template
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......