Pre-run for determining the control limits
Measurement device monitoring for the Fischerscope X-RAY
FISCHERSCOPE
®
X-RAY
123
Mean value x..
7.449
μ
m 7.681
μ
m
Standard deviation s
I
0.125
μ
m 0.133
μ
m
Confidence mean value
0.051
μ
m 0.066
μ
m
Range R(x.)
0.321
μ
m 0.327
μ
m
Lower control limit (x)
7.26
μ
m 7.46
μ
m
Upper control limit (x)
7.64
μ
m 7.90
μ
m
Upper control limit (s)
0.300
μ
m 0.338
μ
m
Number of groups
10 10
Measurement time
20 s
All printed values are available in the print form of WinFTM
PDM as
variables (including the control charts). They can re-calculate the
printed control limits by using the original values (mean value x. and
standard deviations s) and the equations specified in the previous
section. Please note that the pre-run has been concluded within a
short period (here: 17 minutes) in order to ensure the stability of the
instrument during the pre-run.
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......