Programming Coordinates for Automatic Measurements (XDLM 237)
17
Programming Coordinates for Automatic Measurements
(XDLM 237)
The instrument XDLM 237 is equipped with a motor-driven measuring stage. You can program co-
ordinates for measurement points in order to let the instrument measure automatically at various
locations on the specimen. In this manner, you can define, for example, 100 measurement points
and let the measurement run automatically over night at these measurement points.
The following table describes the various programming options for the coordinates and references
the respective chapters.
Arrangement
Function
Chapter
Individual points
You are programming any desired points.
Lines
You define the beginning and end point of a line
and the number of points on this line.
Array
You define two corner points of an array and the
number of points within the array (columns and
rows). In addition, you define the order in which
the points are visited.
Same programming
on multiple parts
( pattern)
You define a single point and any desired num-
ber of points in relation to this single point. You
can repeat this arrangement of points (pattern)
any number of times. In this manner, you can
travel to the same coordinate sequence on differ-
ent parts or at different locations.
Circular coordinates
You define a center of a circle, the number of
measurement points on the circumference of the
circle and the starting point.
Datum points
In addition to the measurement points (individual
points, lines, arrays or circular coordinates), you
can also program one or two datum points. No
measurements are made at these datum points;
rather, the datum points are used for the orienta-
tion of the specimen. You can then change
parts, or place parts anew and ensure that al-
ways the correct coordinates are visited, even if
the specimen is not positioned 100% at the
same place.
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......