FISCHERSCOPE
®
X-RAY
165
Measurement Data Export
Table 20-2: Explanation of the controls of the Exporting data window
Control
Explanation
Online-Export ON
The measurement data is exported continuously when measurement data
is acquired by the instrument.
The reading is exported directly after a measurement is made.
The block statistics are exported upon block closure.
With product selection, the header is exported.
Template and No.
The name and the number of the export template are displayed if an ex-
port template is assigned to the current Product.
►
To assign an export template, click
Select…
.
Deletes the assignment of an export template to the product.
Export destination
append to file
Data is successively written to a file.
Complete file-
handshake
(NET_EXPT.END)
WinFTM waits until the handshake file
NET_EXPT.END
is deleted. This
control box should be enabled only, if the technology of the target com-
puter is capable of handling it.
overwrite file
The export file contains only the last exported values, i. e. the old data
will be overwritten.
Destination file
Displays the destination file, if already assigned.
►
To assign a destination file for the data export, click
Select…
.
Filename variable
By clicking the list box you can select a rule, how the file name of the ex-
ported data will be created.
send to serial port
Data is sent directly to the RS232 interface.
The parameters of the RS232 interface and of the WinFTM software of
the receiving instrument must be the same. Otherwise the receiving instru-
ment will not be able to read the data exported via the RS232 interface.
►
To set the parameters of the RS232 interface select
Evaluation > Export > Define export interface…
.
to OLE-Automation
WinFTM can be remote controlled via OLE and can transfer data via
OLE. A Visual-Basic (Version 6) program example is included on the Win-
FTM installation CD in the directory
Ser-
vice\OLE_Import_Sample_VisualBasic_6.0\AutomationDemo
. This
sample program can be used as a template for the customers own remote
control software.
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......