Product Administration
Backing up product or Def.MA files
60
FISCHERSCOPE
®
X-RAY
3. In the WinFTM main window select
Product > Copy to/from File…
.
The window
Copy from/to file
appears.
4. Select the tab
File >>> Def.MA
.
5. Click
Select
.
The window
Select file
appears.
6. Select the Def.MA file and click
Select
.
The selected path appears in the field
File
.
7. Click
Start copy
.
The file will be copied. After this the report
The Def.MA was copied
successfully
appears.
8. Click
OK
.
9. To close the window, click
cancel
.
What you can do next
You can now create a new product based on the imported Def.MA.
8.12 Backing up product or Def.MA files
Backing up product or
Def.MA files
Product or Def.MA files can be backed up by copying them to a
different folder of the hard disk or of the network environment or
onto a diskette. Product files can be backed up with or without data.
Select the menu command “Product > Copy from / to file” to start
the backup procedure.
8.13 Reference samples
Reference samples
Reference samples (working standards) are specimen with a clearly
defined reference area with a known coating thickness or
composition. Reference samples may be finished parts from one’s
own or someone else’s production that have been measured with a
well and properly calibrated instrument.
Reference samples can be used, for example, to monitor the
measurement devices.
The coating thickness or the composition should be uniform across
the reference area. The reference sample should also exhibit the
same properties (substrate material, composition of coating or
substrate material, etc.) as the specimen for which the product file
has been set up.
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......