FISCHERSCOPE
®
X-RAY
127
Def.MA
16
Def.MA
If you have activated the SUPER software, you can create a Def.MA by yourself.
The usage of the SUPER software requires deep knowledge of the physics of x-ray fluorescence and
the functionality of the instrument.
Alternatively, you can request a new measurement application from FISCHER. FISCHER offers a
professional service for the creation of customer’s Def.MA.
16.1 The Def.MA Modes
When you create a new Def.MA, you can choose between different modes. The mode determines
the measurement method and the presentation of the measurement readings.
Table 16-1: Def.MA Modes
Mode
Application
Material analysis
For the analysis of bulk material (massive specimens).
Analysis with bal-
ance
For the analysis of bulk material (massive specimens), if the specimen can-
not be defined completely by 100 percent.
Typical applications: recycling, RoHS, soil specimens, biological specimens
Thickness mode
For coating thickness measurements. The measurement results for the coat-
ing thickness are displayed in the typical units of length such as μm. The
analysis results are shown in percent units.
The density values of the elements are taken from internal tables. You do
have the option to change these values.
Mass per unit
area (relative)
For coating thickness measurements. The overall area mass is displayed as
well as the percentage portion of the individual elements.
Mass per unit
area (absolute)
For coating thickness measurements. The overall area mass is displayed as
well as the individual area masses of the respective elements.
Solution analysis
You fill the measuring cell with the solution to be analyzed, see Chap. 19.2
on Page 156.
ROI-Mode
ROI = Region of interest
Only a defined portion of the measures spectra is evaluated. You define
one or more regions (ROI) of energy channels. The portion of the spectra
that is located in the defined region is evaluated only.
Counting rates are issued, the fundamental parameter method is not used.
Used for special applications, where the fundamental parameter method is
not applicable.
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......