180
FISCHERSCOPE
®
X-RAY
Pattern Recognition (XDLM 237)
9. Define the measuring position by clicking on the measuring position with the right mouse
button. The measuring position can be inside or outside of the search item rectangle.
10.Set the parameters according to the following table:
11.Click
Show search area
.
Another hatched rectangle appears in the pattern recognition window. This rectangle marks the
search area, i.e. the area of the image in which the image details are to be searched for.
Figure 21.5: Pattern recognition window: The hatched rectangle shows the search area
Parameter
Value
With progr. XY-Pos.
Activated
On measure, find automatically
Activated
Algorithms area
Mode
2 = Norm.corr
Center weight
0.0 = no
Error check
5% error
Max. rotation
+/- 110 degree
Search area