160
FISCHERSCOPE
®
X-RAY
Solution Analysis
7. Click
Start copy
.
The file will be copied. After this the report
The Def.MA was copied successfully
appears.
8. Click
OK
.
9. To close the window, click
cancel
.
What you can do next
You can now create a new product based on the imported Def.MA.
19.5.3 Setting up a Product File for the Solution Analysis
Before you start
Before you can set up a product for the solution analysis, a Def.MA file for your solution analysis
must be present on the computer.
Procedure
1. In the WinFTM main menu select
Product > New…
.
The window
Select an application
appears.
2. Click
New
.
The window
Select a Def.MA
appears.
3. Select the Def.MA and click
OK
.
The window
Input of the application name
appears.
4. Enter in the field
Name
an identifier for the new measurement application and click
OK
.
If your instrument is equipped with changeable apertures (collimators), the window
Select
collim. + meas. dist.
appears. Select the standard collimator No. 2 and click
OK
.
The window
Input Calibr. Std. Set
appears.
5. Do not enter anything in this window and click
OK
.
6. If your are prompted to measure
scatt
, do not perform a measurement, but perform the
following steps instead:
a In the WinFTM main menu select
Tools > Load Spectrum and Evaluate…
.
b If the window
Open
appears, select the file
scatt
and click
Open
.
The window
End of calibration. Accept?
appears.
7. Click
Yes
.
8. The window
Create new product
appears.
9. Enter a name, directory and number for the new product and click
OK
.
10.To check the measurement time and the number of measurements click
Info
.
The window
Modify product
appears.
11.Click the tab
Product
.
12.Check the values for the measurement time and the number of measurements. Change the
values, if necessary.
The longer the measurement time, the better the repeatability precision. Do not select a meas-
uring time that is too short; 30 or 60 seconds is a typical value.
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......