Instrument Overview
2.4
Functional Principle of the Instrument
The following figure shows the principle structure of the instrument:
Figure 2.2: Functional principle of the instrument
Functional principle
1. The x-ray tube generates the primary x-radiation (primary radiation). The electrically heated
cathode emits electrons. Accelerated by the applied high voltage to very high speeds, the elec-
trons bombard the anode material. This generates the primary x-radiation.
2. The primary filter optimizes the energy distribution of the primary x-radiation.
3. The shutter serves as a safety device. It is opened during a measurement only. In case of a
malfunction or when opening the measurement chamber, it will be closed automatically. This
ensures, that the user is never exposed to x-radiation.
X-ray tube
Cathode
Anode
Primary x-radiation
Aperture (Collimator)
Coating layer
Base material
Detector
Shutter
Mirror
Optics
Video camera
Spectrum
WinFTM
main window
X-ray fluorescence radiation
Primary filter
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......