FISCHERSCOPE
®
X-RAY
151
Calibration, Base Correction and Reference Measurement
After the last measurement,
the window
Input of the cali-
bration standard set
will
appear.
14.Ensure that the following conditions are met:
The values in columns
(3) Corr.
(corrected values) are consistent with the values in columns
(1) Set
(nominal values).
The values in columns
(2) Theor.
(standard-free values) deviate from the values in columns
(1) Set
(nominal values) by no more than 20 %.
The values in columns
(2) Theor.
(standard-free values) should fit logically, i.e., a higher
value in column
(2) Theor.
(standard-free values) belongs to a higher value in column
(1)
Set
(nominal value).
15.if the conditions are not met, check your settings and repeat the calibration beginning with
16.When the conditions are met, click
OK
.
The calibration for this measuring application will be activated.
17.To verify the calibration, select
Product > Measure Cal. Standards
.
The name of the calibration
standard set will appear in
the field Product name with
a red background.
18.Measure the calibration standards as prompted. Measure each standard several times.
19.Verify that the nominal value of the standard is within the range x (mean value) ±2 s (standard
deviation of the re-measurement).
The following table shows some examples:
1
2
3
1
2
3
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......