Features of WinFTM BASIC and LIGHT
2.12 Presentation of Readings
You can control, how the measurement readings are presented.
2.13 Calibrating and Monitoring of the Measuring Equipment
Display mode
Description
Thickness mode
The measurement result shows:
Coating thickness readings in the common length dimensions
e.g. in mils
Analysis results in percent
The densities are taken from internal tables, but you do have the
option of changing the values.
Absolute area mass mode
The measurement result shows:
Total area mass
Individual area masses of the respective elements, e.g., in g/
m² or g/l.
No densities needed.
Relative area mass mode
The measurement result shows:
Total area mass
Relative area masses of the respective elements in %.
No densities needed.
Component mode
Used for the analysis of specimens, which are not only composed
of pure elements, but also contain chemical compounds (compo-
nents).
Main application fields: Oxides, metallic alloys, …
Function
Description
Calibration of new or exist-
ing products
The obtained calibration data are stored.
Standard-free calibration
You can calibrate without calibration standards by means of the
fundamental parameter method.
Standard-based calibration
with the use of up to 64
standards
You can correct the standard-free calibration by calibrating with
calibration standards. Thus, the accuracy of the analysis and thick-
ness measurements is improved.
Matrix-specific calibration standards and pure element calibration
standards are available. Up to 64 calibration standards can be
used for each measurement application.
Summary of Contents for FISCHERSCOPE X-RAY XDLM 231
Page 8: ...6 FISCHERSCOPE X RAY Table of Contents...
Page 14: ...12 FISCHERSCOPE X RAY Safety Information...
Page 30: ...28 FISCHERSCOPE X RAY Set up...
Page 36: ...34 FISCHERSCOPE X RAY WinFTM File Structure...
Page 52: ...50 FISCHERSCOPE X RAY User Interface of the WinFTM Software...
Page 134: ...132 FISCHERSCOPE X RAY Def MA...
Page 146: ...144 FISCHERSCOPE X RAY Programming Coordinates for Automatic Measurements XDLM 237...
Page 186: ...184 FISCHERSCOPE X RAY Pattern Recognition XDLM 237...
Page 192: ...190 FISCHERSCOPE X RAY Cleaning and Maintenance...
Page 202: ...200 FISCHERSCOPE X RAY Addendum...
Page 228: ...226 FISCHERSCOPE X RAY WinFTM SUPER For the Experienced X RAY User...
Page 229: ...FISCHERSCOPE X RAY 227...
Page 232: ......